User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Power Designs 2020B:Sub 1 DATE: 12-Mar-96 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 11 NUMBER OF LINES: 87 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK+ K X 1.002 ASK- P F 1.003 CALL 2020B Sub 5 1.004 ACC 200 .100mV 7P% 0.5P/ 1.005 MEME 1.006 MEMC mV 0.1% 0.5U 1mVÿRange #! Test Tol 0.5001, Sys Tol 0.0001007, TUR 4966.236 (>= 4.00). 2.001 DISP Set the UUT front panel controls as follows: 2.001 DISP 0-1mV vernier ..................................... .2mV 2.002 CALL 2020B Sub 5 2.003 ACC 200 .200mV 7P% 0.5P/ 2.004 MEME 2.005 MEMC mV 0.1% 0.5U 1mVÿRange #! Test Tol 0.5002, Sys Tol 0.0001014, TUR 4932.939 (>= 4.00). 3.001 DISP Set the UUT front panel controls as follows: 3.001 DISP 0-1mV vernier ..................................... .3mV 3.002 CALL 2020B Sub 5 3.003 ACC 200 .300mV 7P% 0.5P/ 3.004 MEME 3.005 MEMC mV 0.1% 0.5U 1mVÿRange #! Test Tol 0.5003, Sys Tol 0.0001021, TUR 4900.098 (>= 4.00). 4.001 DISP Set the UUT front panel controls as follows: 4.001 DISP 0-1mV vernier ..................................... .4mV 4.002 CALL 2020B Sub 5 4.003 ACC 200 .400mV 7P% 0.5P/ 4.004 MEME 4.005 MEMC mV 0.1% 0.5U 1mVÿRange #! Test Tol 0.5004, Sys Tol 0.0001028, TUR 4867.704 (>= 4.00). 5.001 DISP Set the UUT front panel controls as follows: 5.001 DISP 0-1mV vernier ..................................... .5mV 5.002 CALL 2020B Sub 5 5.003 ACC 200 .500mV 7P% 0.5P/ 5.004 MEME 5.005 MEMC mV 0.1% 0.5U 1mVÿRange #! Test Tol 0.5005, Sys Tol 0.0001035, TUR 4835.749 (>= 4.00). 6.001 DISP Set the UUT front panel controls as follows: 6.001 DISP 0-1mV vernier ..................................... .6mV 6.002 CALL 2020B Sub 5 6.003 ACC 200 .600mV 7P% 0.5P/ 6.004 MEME 6.005 MEMC mV 0.1% 0.5U 1mVÿRange #! Test Tol 0.5006, Sys Tol 0.0001042, TUR 4804.223 (>= 4.00). 7.001 DISP Set the UUT front panel controls as follows: 7.001 DISP 0-1mV vernier ..................................... .7mV 7.002 CALL 2020B Sub 5 7.003 ACC 200 .700mV 7P% 0.5P/ 7.004 MEME 7.005 MEMC mV 0.1% 0.5U 1mVÿRange #! Test Tol 0.5007, Sys Tol 0.0001049, TUR 4773.117 (>= 4.00). 8.001 DISP Set the UUT front panel controls as follows: 8.001 DISP 0-1mV vernier ..................................... .8mV 8.002 CALL 2020B Sub 5 8.003 ACC 200 .800mV 7P% 0.5P/ 8.004 MEME 8.005 MEMC mV 0.1% 0.5U 1mVÿRange #! Test Tol 0.5008, Sys Tol 0.0001056, TUR 4742.424 (>= 4.00). 9.001 DISP Set the UUT front panel controls as follows: 9.001 DISP 0-1mV vernier ..................................... .9mV 9.002 CALL 2020B Sub 5 9.003 ACC 200 .900mV 7P% 0.5P/ 9.004 MEME 9.005 MEMC mV 0.1% 0.5U 1mVÿRange #! Test Tol 0.5009, Sys Tol 0.0001063, TUR 4712.135 (>= 4.00). 10.001 DISP Set the UUT front panel controls as follows: 10.001 DISP 0-1mV vernier ...................................... 1mV 10.002 CALL 2020B Sub 5 10.003 ACC 200 1.000mV 7P% 0.5P/ 10.004 MEME 10.005 MEMC mV 0.1% 0.5U 1mVÿRange #! Test Tol 0.501, Sys Tol 0.000107, TUR 4682.243 (>= 4.00). 11.001 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.